Index > News Center > Enterprise dynamics > What are the precautions for using the ultrapure water machine again after the holidays?
Release time:2026-11-20
Views:1Risk warning: After the holidays, the interior of the ultrapure water machine, which has been left idle for a long time, may harbor microorganisms, form bacterial films, or accumulate stagnant water. Using it directly can affect water quality and experimental data. The core principle is to thoroughly flush and drain the stagnant water first, and only use it after confirming that the water quality meets standards.
1. Inspection before startup
Check the inlet water pressure: 0.2-0.4 MPa, ensure the pipeline is free of bends, the joints are not leaking, and the drainage pipeline is unobstructed.
First, open the water inlet valve and allow water to flow through for 10-20 seconds to exhaust any air. Then, connect the device to power. It is prohibited to operate the device without water to avoid damaging the booster pump and RO membrane due to dry running.
Check the pre-treatment filter elements (PP cotton, activated carbon), and replace them if they appear yellowed or blackened; if the machine has been shut down for more than 1 month, it is recommended to check the condition of the consumables.
Storage tank: If the tank is not emptied during the holidays, all stored water must be drained before refilling. Observe for any unusual odor or slimy bacterial film, as these indicate a need to clean and disinfect the tank.
II. Startup flushing operation (🔑 Key steps)
The equipment completes the self-inspection after being powered on and then performs a complete machine flushing.
RO pure water port: let the water run for 5-10 minutes to drain the stagnant water in the pipeline, and observe the stability of RO conductivity;
UP ultra-pure water nozzle: let water run for 5-10 minutes;
After the water tank is filled with water, it is recommended to drain it completely and then refill it to complete the whole machine cycle activation.
If the machine is shut down for more than 7 days: the flushing time should be extended to 15-20 minutes; in high-temperature environments during summer, microorganisms are more likely to proliferate, so the flushing time should be appropriately extended.
III. Water quality confirmation: it can only be used after meeting the standards
We must wait for the parameters to stabilize before conducting experiments with water, rather than immediately starting the machine and taking water for testing
RO pure water: stable conductivity;
UP ultrapure water: The resistivity is stably close to 18.2MΩ·cm (25℃), and the TOC value has stabilized after a decrease;
If the resistivity cannot be increased and the TOC remains high:
Continue to rinse in a circular manner;
If the issue persists after rinsing, investigate whether the polishing purification column and terminal filter are contaminated, and replace them if necessary;
If there is an odor or a slippery feeling, the entire machine's piping needs to be disinfected.
Tip: Use ultrapure water immediately after taking it out. If left for a few minutes, it will be contaminated by CO₂ in the air, resulting in a decrease in resistivity.
IV. Usage Grading Suggestions
At the beginning of the restart, RO pure water should be used preferentially for cleaning utensils;
After the resistivity and TOC have stabilized and met the standards, UP ultrapure water can be used for precision experiments such as HPLC, ICP, and PCR to avoid interference from bacterial films and endotoxins on the experimental results.
V. Common fault prompts
The conductivity of the startup solution cannot be increased: This is often due to stagnant water in the pipeline or contamination of the polishing resin. Extend the cycle of flushing; if flushing is ineffective, replace the ultrapure column.
High RO conductivity: It remains high after flushing. Check the RO membrane. Long holidays and downtime can easily lead to microbial contamination of the RO membrane.
The water has a fishy smell and a slippery feeling: microbial growth in the system requires disinfection of the entire machine, and replacement of the terminal filter if necessary
About us
Company profileHonorary qualificationEnterprise dynamicsTechnical articlesProduct display
Laboratory and medical sewage treatment system Water system of laboratory science and Education Institute Water system for pharmaceutical industryColumn navigation
Product Center Project case Video Center Service supportContact us
Online MessageContact us
copyright © 2022 Nanjing Yipu Yida Technology Development Co., Ltd Al Rights Reseved Record number:苏ICP备19073731号-1 Sitemap.xml Manage login technical support:Chemical Industry Instrument Website