Index > News Center > Enterprise dynamics > The difference between pure water and ultrapure water
Release time:2026-11-11
Views:6The core difference between pure water and ultrapure water lies in purity. Ultrapure water is obtained by subjecting pure water to more complex and stringent purification processes, which remove almost all impurities from the water, resulting in a purity level far exceeding that of ordinary pure water.
The difference between pure water (RO Grade III water) and ultrapure water (UP / Grade I water)
According to the GB/T 6682-2008 standard for laboratory water, ultrapure water is obtained by further purification of pure water.
Core indicators (25°C)
The following table can help you better understand their specific differences in a more intuitive way:
project
Pure water (RO, Grade III)
Ultra-pure water (UP, Grade 1 water)
resistivity
0.2‑1 MΩ·cm
18.2 MΩ・cm (theoretical limit)
conductivity
1‑10 μS/cm
≈0.055 μS/cm
TOC Total Organic Carbon
≤50‑200 ppb
≤5‑10 ppb
Impurity grade
ppm level
At ppb or even ppt levels, the ions are almost completely removed
Microorganism
≤10‑100 CFU/mL
<0.1 CFU/mL, capable of removing endotoxins and nucleases
preparation process
Pure water (RO): Pretreatment + Reverse Osmosis RO; partially combined with EDI to remove most salts, colloids, and particulates, with a simple process.
Ultrapure water (UP): Using pure water as the feedstock, followed by: 185nm UV TOC degradation + polishing mixed bed + ultrafiltration for pyrogen removal + 0.22μm terminal microfiltration, achieving multi-stage deep purification.
Ultra-pure water is highly susceptible to contamination by CO₂ in the air and must be used immediately after preparation, as prolonged storage is not advisable. The resistivity will significantly drop within minutes after dispensing.
Usage differentiation
Pure Water RO
Rough washing and final washing of glassware
Autoclave, water bath, ultrasonic cleaning
General Chemistry Experiments, Preparing General Buffer Solutions
As the water source for the ultra-pure water machine
Ultra-pure water UP
Precision instruments such as HPLC, LC-MS, ICP-MS, and atomic absorption
PCR, cell culture, molecular biology experiments
Trace and microanalysis, preparation of standard blank solutions
Popular summary
Pure water: Most impurities are removed, leaving only trace amounts of ions and organic matter, suitable for routine experimental cleaning and general preparation
Translate directly and output the translation without explanation, ensuring the meaning is coherent.
Ultra-pure water: Almost all ions, organic matter, particles, and bacteria are removed; used for high-precision analytical experiments; higher cost, must be prepared and used immediately
Equipment Operation Tips (Pure Water Machine)
RO outlet delivers pure water; UP outlet delivers ultrapure water;
Do not use ultrapure water to wash bottles, as it is wasteful; prefer RO purified water for cleaning;
If the ultrapure water system is not used for an extended period, it should be circulated regularly to prevent water quality degradation and resin contamination.
About us
Company profileHonorary qualificationEnterprise dynamicsTechnical articlesProduct display
Laboratory and medical sewage treatment system Water system of laboratory science and Education Institute Water system for pharmaceutical industryColumn navigation
Product Center Project case Video Center Service supportContact us
Online MessageContact us
copyright © 2022 Nanjing Yipu Yida Technology Development Co., Ltd Al Rights Reseved Record number:苏ICP备19073731号-1 Sitemap.xml Manage login technical support:Chemical Industry Instrument Website