Index > News Center > Enterprise dynamics > Quick! Daily Maintenance and Regular Care for Pure Water Machines!
Release time:2026-11-03
Views:12Maintenance is categorized into daily/weekly routine, monthly maintenance, quarterly/semi-annual maintenance, and annual maintenance, along with common alerts and key points for consumable replacement, suitable for writing operational SOPs and equipment logs.
1. Daily Maintenance (Daily and Weekly, performed by operators)
daily
Before starting, check the inlet water pressure: the inlet pressure should be 0.1‑0.4MPa. Address any low/high pressure promptly. Inspect for leaks or water seepage at pipe connections.
Check the device display: Inlet TDS, RO Product TDS, Resistivity (Ultra-pure Water), Temperature, Alarm Code, and confirm the water quality is within the specified range.
After water collection: It is recommended to drain the ultra-pure water side for 30-60 seconds to remove stagnant water in the pipeline and ensure water quality.
Observe whether the wastewater discharge is normal, with no blockages or abnormal noises.
weekly
Clean the exterior surfaces of the equipment, wipe the cabinets and displays, and avoid direct water rinsing of electrical components.
Inspect the appearance of the pretreatment filter (PP cotton) for discoloration or clogging; check the inlet water filter for sand blockage.
Inspect the water tanks: Check whether the liquid level float ball in the raw water tank and pure water tank is functioning normally; examine whether the appearance of the tank vent filter is blackened or dust-clogged.
The system must run at least once a week to ensure water circulation in the pipelines and prevent microbial growth. For prolonged periods of inactivity (exceeding 3 days), the system should be drained and powered off.
⚠️ If the system has been out of service for more than 7 days, it cannot be restarted directly for water extraction. Flushing and disinfection must be performed to prevent microbial contamination exceeding standards.
II. Monthly Maintenance
Inspect the pre-filtration elements: PP cotton and activated carbon, checking for pressure differential and water output. If the pressure differential increases or water output decreases, replace them in advance rather than waiting until the scheduled replacement date.
Inspect the pipelines, valves, and joints to ensure there are no leaks; tighten any loose connections.
Inspect the water tank air vent filter. Replace it immediately if it is dusty, damp, or discolored to prevent bacteria from the air from entering the pure water tank.
Check the operating pressure and recovery rate of the RO membrane, record the operational parameters (inlet pressure, product water pressure, TDS, resistivity), and establish an equipment operation log.
Inspect the resistivity electrodes: Check if the probes have any scaling. If there is minor scaling, remove and clean them to ensure accurate measurement readings.
compressed activated carbon (in poor water quality environments, the replacement interval can be shortened to 2 months).
Clean the inner walls of the purified water tank (if visible deposits or slippery mucus are present, perform tank disinfection).
Check the operating noise and pressure stability of the booster pump; inspect whether the solenoid valve functions properly.
every 6 months
Chemical Cleaning of Reverse Osmosis RO Membranes
When the RO desalination rate significantly declines and the water production drops by more than 15%, cleaning is mandatory; the interval can be extended if the condition remains good.
Monitor the resistivity of the product water from the ultra-pure water purification column (DI column). Replace it immediately if it fails to meet the standards, rather than waiting until it completely fails.
Whole machine pipeline + Pure water tank disinfection
Inhibiting bacteria and microorganisms is a point often overlooked by laboratory pure water machines; microbial contamination can directly affect experiments.
Inspect the wastewater discharge pipeline to prevent backflow caused by blockages.
IV. Annual Maintenance
1. Deep maintenance of RO membranes: Clean or evaluate whether the RO membrane elements need replacement.
2. Replace all pre-filters, inspect the high-pressure pump, and the pressure regulating valve.
3. Verify the resistivity/conductivity measurement electrodes to ensure accurate water quality readings.
4. Inspect all pipelines for aging, replace hardened or cracked pipelines promptly; check the sealing rings.
5. Maintain a complete record of the entire equipment operation report.
V. Reference Cycle for Key Consumables Replacement (For Reference Only; Shorten if Inlet Water Quality is Poor)
Consumables
Regular replacement cycle
```Criteria for Failure Judgment```
PP/PC
Filter element
3-6 months
Reduced water output, high pressure differential, and darkened filter element
Particle/sintering
activated charcoal
3-6 months
Excessive free chlorine and foul odor in the effluent
RO reverse osmosis membrane
12 to 15 months
The desalination rate decreased by more than 15%, and the water production significantly declined
Purification column/ultrapurification column
Depending on water usage
If the resistivity does not reach 18.2MΩ・cm at 25℃, replace immediately
6. Important Notes
1. Water quality conditions determine lifespan: High sediment and chlorine levels in the influent will shorten the lifespan of all consumables. Do not exceed the service life of consumables by force.
2. Avoid prolonged shutdown without treatment; for short-term shutdowns (3-7 days), perform short-cycle flushing daily; for shutdowns exceeding 7 days, empty the water tank and implement protective measures.
3. Disinfectant agents must be specifically prepared using purified water machine disinfectants. Household 84 disinfectant is strictly prohibited from being directly introduced into the system, as it may damage the RO membrane and purification column.
4. Maintain a log record for each maintenance and consumable replacement, including the replacement time, consumable model, and water quality parameters.
About us
Company profileHonorary qualificationEnterprise dynamicsTechnical articlesProduct display
Laboratory and medical sewage treatment system Water system of laboratory science and Education Institute Water system for pharmaceutical industryColumn navigation
Product Center Project case Video Center Service supportContact us
Online MessageContact us
copyright © 2022 Nanjing Yipu Yida Technology Development Co., Ltd Al Rights Reseved Record number:苏ICP备19073731号-1 Sitemap.xml Manage login technical support:Chemical Industry Instrument Website